SeqBench

C1V1 = C2V2, and When It Stops Being Usable

7 min read · Updated November 11, 2026

C1V1 = C2V2 answers a question that is usually not the real one. It tells you what volume of stock contains the amount you need; it says nothing about whether you can pipette that volume, and at high dilution ratios the answer is no.

The useful version of the equation therefore has a second half: if the stock volume falls below about a microlitre, split the dilution into steps. The arithmetic for that is the only part of this that people get wrong.

The equation, and the one rearrangement worth memorising

C1V1 = C2V2, where C1 and V1 are the stock and C2 and V2 the final. Solving for the stock volume: V1 = C2 × V2 ÷ C1. That is the whole calculation, and it works in any consistent units — nM and µL, ng/µL and mL, as long as both concentrations are in the same unit.

The reason it is worth writing down rather than doing in your head is unit mixing. A stock labelled 100 µM and a target written 200 nM are the same kind of quantity three powers of ten apart, and the usual error is to enter 200 as though it were µM and get a plan that asks for fifty times too much stock.

Worked: 100 µM to 200 nM in 50 µL

V1 = 200 nM × 50 µL ÷ 100,000 nM = 0.1 µL. The arithmetic is right and the instruction is not deliverable: a single-channel pipette does not accurately deliver a tenth of a microlitre, and on a 500-fold dilution the error in that transfer is most of the final concentration.

The fix is two steps rather than one. Split 500-fold into two equal steps of √500 ≈ 22.4: take 2.236 µL of stock into 47.76 µL of diluent, which gives 4,472 nM, then take 2.236 µL of that into 47.76 µL again, which gives 200 nM. Both transfers are well inside the pipette's accurate range, and the two steps multiply to 500.

Note what did not change: the final volume is still 50 µL each time, so the diluent volume follows from the transfer rather than being chosen first.

Worked: 20,000-fold, where the step count comes from

A primer working solution at 5 nM from a 100 µM stock in 20 µL is a 20,000-fold dilution. The direct route would ask for 0.001 µL of stock.

How many steps? One step can dilute by at most V_final ÷ V_min, because the transfer is V_final ÷ fold. With a 20 µL final volume and a 1 µL floor, the largest single step is 20-fold, so 20,000-fold needs ceil(ln 20000 ÷ ln 20) = 4 steps. Four equal steps of 20,000^(1/4) = 11.89-fold each: 1.682 µL into 18.32 µL, four times, landing on 8,409 → 707 → 59.5 → 5 nM.

The step count is the part that gets skipped. Splitting a 20,000-fold dilution into two steps looks reasonable and does not work: two equal steps are √20,000 ≈ 141-fold each, so in a 20 µL volume each transfer is 0.14 µL — better than the single step's 0.001 µL, but still seven times below the 1 µL floor, and now there are two error-carrying transfers instead of one. Splitting has to go far enough to clear the floor, which is what the logarithm is for.

Worked: when the units are not the same on both sides

The interesting case is a mass concentration on one side and a molar one on the other, which is what happens whenever the stock is what the spectrophotometer said and the target is what the protocol asked for. A 3 kb plasmid at 500 ng/µL is 256.4 nM, because 3,000 bp × 650 g/mol/bp = 1,950,000 g/mol. To reach 10 nM in 100 µL: V1 = 10 × 100 ÷ 256.4 = 3.9 µL of stock into 96.1 µL.

Without the molar mass there is no way to cross between those units, and estimating it from length rather than computing it from the sequence is where the few-percent errors enter. For a plasmid the 650 g/mol per bp average is good enough; for a short oligo it is not, and the exact composition is worth using.

Three ways the plan is wrong even when the arithmetic is right

  • Diluting into a volume that is then aliquoted. Each aliquot carries the dilution error again; make the dilution once at a volume big enough to freeze in single-use aliquots.
  • Reusing a tip for the diluent and the stock. At high fold-dilution the carryover is a meaningful fraction of the stock that reaches the tube.
  • Assuming a dilution is stable. Oligo working stocks at nanomolar concentration adsorb to plastic and degrade faster than the frozen stock, which is why the working solution is made fresh rather than stored.

FAQ

What is the C1V1 = C2V2 formula?

C1 × V1 = C2 × V2 for a dilution, so the stock volume is V1 = C2 × V2 ÷ C1. Both concentrations must be in the same unit, which is where most errors enter — 100 µM against 200 nM is 100,000 against 200.

How do I make a dilution that needs less than 1 µL of stock?

Split it into serial steps. A single step can dilute at most V_final ÷ V_min, so with a 50 µL final volume and a 1 µL floor the largest step is 50-fold; take the number of steps as the logarithm of the total fold divided by the logarithm of the per-step limit.

Why is a serial dilution more accurate than one big dilution?

Because accuracy is limited by the smallest volume you pipette. One 500-fold dilution in 50 µL asks for 0.1 µL; two 22.4-fold steps ask for 2.24 µL each. The error of two good transfers is far smaller than the error of one bad one.

How do I convert ng/µL to nM?

Divide the mass concentration by the molar mass: nM = (ng/µL) ÷ (g/mol) × 10⁶. For double-stranded DNA the molar mass is about 650 g/mol per base pair, so a 3 kb plasmid is about 1.95 × 10⁶ g/mol and 500 ng/µL is 256 nM.

Related references

Related tools

Related guides

Helpful?